System Commands 5
WA VEFORM TRANSFER
DESCRIPTION
COMMAND SYNTAX
.QUERY SYNTAX
Response format
STORE_SETUP, STST
Command/Query
Oscilloscopes fitted with the MC01 Option
The STORE_SETUP command controls the way in which traces
will be stored on the memory card. Traces may be enabled (FILE)
or disabled (DIS) for storage. This applies to autostoring or to the
CARD STORE STST command. Traces may be autostored to the
memory card after each acquisition until the card becomes full
(FILL), or forever (WRAP), replacing the oldest traces by
ones. They may be stored keeping the full precision of the data
(WORD), or with reduced precision (BYTE, most significant bits
only) to conserve space on the card.
The STORESETUP? query returns the current mode of opera-
tion of Autostore, the current data size, and the current trace
selection.
SToreSeTup [<trace>,<enable> ..... <trace>,<enable>]
[,AUTO,<mode>] [,SIZE,<data size>]
<trace> := {EA, EB, MC, MD, FE, FF, C1, C2, C3:1:, C4~;}
<enable> := {FILE, DIS}
<mode> := {OFF, WRAP, FILL}
<clata size := {WORD, BYTE}
Note 1" Parameters are grouped in pairs. The first one names the
variable to be modified and the second one gives the new value to
be assigned. Pairs may be given in any order and may be restricted
to those variables which have to be changed.
Note 2: The STORE_SETUP command only modifies the current
setup for storing traces to the memory card. Any parameters not
given in a particular STORE_SETUP command will not be af-
fected, i.e. they will stay at their previous settings.
STore SeTup?
STore_SeTup <trace>,<enable> ..... <trace>,<enable>,AUTO,
<mode>,SIZE,<size>
4-channel oscilloscopes only