System Commands 5
ACQUISITION
DESCRIPTION
COMMAND SYNTAX
QUERY SYNTAX
Response format
EXAMPLE (GPIB)
VOLT_DIV, VDIV
Command/Query
The VOLT DIV command sets the vertical sensitivity in Volts/div.
The VAB b~ (bit 2) in the STB register (Table 8, page 137 ) is set
an out-of-range value is entered.
Note: The probe attenuation factor is not taken into account for
adjusting vertical sensitivity.
The VOLT_DIV? query returns the vertical sensitivity of the speci-
fied channel.
<channel>:Volt_DIV <v_gain>
<channel> := {C1, C2, C35, C4,}
<v_.gain> :=
9410114130 All others
1 mV to 2.5 V
5.0 mV to 2.5 V
Note: The suffix V is optional.
<channel>:Volt DIV?
<channel>:Volt_DIV <v_gain>
The following command sets the vertical sensitivity of channel 1 to
50 mV/div.
CMD$="C I:VDIV 50MV" :
CALL IBWRT(SCOPE%,CMD$)
:~ 4-channel oscilloscopes only
163