Rev. D Performance Verification 5-25
b. DC 50
with Bandwidth Limiter On
!
Recall
xxxP024.PNL
or configure the DSO:
Panel Setups :
Recall FROM DEFAULT SETUP
Channels Trace ON
Channel 1
Input Coupling :
DC 50
Global BWL :
20 MHz
Input gain :
100 mV/div.
Input offset :
0 mV
Trigger setup :
Edge
Trigger on :
1
Slope line :
Pos
Mode :
Auto
Time base :
1
µ
µµ
µ
sec/div
.
Channel use :
4
Record up to :
25 k
WAVE PILOT :
MEASURE
Mode :
Custom
Statistics :
Off
Change parameters
On line 1 :
Sdev of 1
On line 2 :
Freq of 1
!
Connect the test equipment as shown in Figure 5-6.
!
Set the generator frequency to
300 kHz
.
!
Adjust the generator signal amplitude to measure
sdev(1)
=
200 mV.
!
Set Time base :
50 nsec/div
.
!
Increase the generator frequency until
sdev(1)
=
140 mV.
(typically 25 MHz)
!
Press
Clear Sweeps
!
When
sdev(1)
=
140 mV,
record Freq(1) in Table 9.
!
Check that the frequency is within the limits specified in Table 9.

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the LeCroy Waverunner2 LT262 Series and is the answer not in the manual?

LeCroy Waverunner2 LT262 Series Specifications

General IconGeneral
BrandLeCroy
ModelWaverunner2 LT262 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals