6.5.5 Power Supply Voltage Problem
AC Line
Voltage 90 to
250V OK ?
Repair as Need.
AC Power
Cord OK ?
Repair Power
Cord.
N
Y
N
Y
Is AC Line
Fuse on Rear
Panel O
p
ened.
N
Y
Is Fan-mortor
Stopped.
N
Y
Is Power
Supply Load
Shorted ?
N
Y
Is Fan-mortor
Broken ?
N
Y
Over Load
Protection.
Replace the
Defective Board
or Unit.
Over Heat
Protection.
Replace Fan-mortor
Assembly.
Cool Down and Take Off AC
Power Cord from Power
Supply Unit over 10
seconds.
Power Supply is
Broken.
Replace Power
Supply Unit.
End
Start
Maintenance 6-9