EasyManuals Logo

LeCroy Waverunner2 Series User Manual

LeCroy Waverunner2 Series
155 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #81 background imageLoading...
Page #81 background image
Rev. D Performance Verification 5-45
5.11.2 Trigger on Pulse Width 100 nsec
a. Pulse Width < 100 nsec
!
Recall
xxxP047.PNL
or configure the DSO as shown in 5.11.1.a
and make the following changes :
Width < 100 nsec :
On
Width > 100 nsec
:Off
Time base :
20 nsec/div.
!
Set the generator frequency to
10 MHz
.
!
Check that the scope Triggers. Record the test result in Table 12.
!
Set Width < 100 nsec
Off
and
Width > 100 nsec
ON
!
Check that the scope
doesn’t trigger :
slow trigger and no flashes in box next
to normal. Record the test result in Table 12.
b. Pulse Width > 100 nsec
!
Set the generator frequency to
4 MHz
.
!
Time base :
50 nsec/div.
!
Set Width < 100 nsec
Off
and
Width > 100 nsec
ON
!
Check that the scope Triggers. Record the test result in Table 12.
!
Set Width < 100 nsec
On
and
Width > 100 nsec
Off
!
Check that the scope
doesn’t trigger :
slow trigger and no flashes in box.
Record the test result in Table 12.