3.2 Technical data
General
Approvals: The instrument-specific approval marks are located at the
rear panel of the instrument, next to the name plate.
Nominal supply voltages: 100 to 120 V ~ +/- 10 %
200 to 240 V ~ +/- 10 %
Nominal frequency: 50 to 60 Hz
Power fuse: 2x T 3.15 A L250 V
Maximum power draw at 100 - 120 V: 4.0 A
Maximum power draw at 200 - 240 V: 2.8 A
Leakage current at 240 V/50 Hz: ca. 2.4 mA
Nominal power: 700 VA
IEC 1010 classification: Protection class 1, pollution degree 2
Overvoltage installation category II
Operating elevation: up to max. 2000 m NN
A-weighted noise level: < 70 dB (A)
IP protection class (IEC 60529) IP20
Climatic conditions for the operation of the instrument:
Temperature: +15 °C to +30 °C
Relative humidity: 20 - 85 % - non-condensing
Climatic conditions for the storage and transport of the packaged instrument:
Storage temperature range: +5 °C to +50 °C
Transport temperature range: -29 °C to +50 °C
Relative humidity: 10 - 85 % - non-condensing
Dimensions and weight:
Dimensions of basic instrument:
Width x depth: 475 x 650 mm
Height with magazine: 895 mm
Dimensions with unload station
connected:
Width x depth: 548 x 650 mm
Height with magazine: 655 mm
Basic instrument empty weight: approx. 28 kg
Weight, packed: approx. 65 kg
Unload station empty weight: approx. 14 kg
Weight, packed: approx. 32 kg
Performance:
Load capacity: up to 6 magazines,
up to 80 cassettes per magazine