MPX G2 Service Manual
4-10
Feel Good Circuit On
21. Press the Yes button 8 times. The display reads: Devel FGood In 1. Turn the knob CCW so the setting
is Out 0 this will disengage the Feel Good circuit.
22. Again apply a 1kHz-sinewave signal Burst at -6dBu (388 mVRMS).
23. Set the Analyzer for a 0dB reference (@ 1kHz).
24. Verify the waveform measurement looks like the following signal in the figure below with the Feel Good
circuit off.