Plug-in control
3.10 Operating menu
3.10.1 Basic Menu
Menu item Description Adjustable value / option Ac-
cess
min. max. stan- Unit
value value dard
Ready Operating display - - - - -
Freq. Setpoint Sets the speed for operation 150 **Hz r/w on
!! Every change is directly written
into the pump’s data storage
and is valid immediately!!
Motor Temp. Motor temperature actual value °Cr
Converter Temp. Temperature of the power actual value °Cr
electronic
Bearing Temp. Temperature of the magnetic actual value °Cr
bearing
Actual Value TMS Temperature of the Temperature actual value °Cr
Management System
T_A Value for Cooling Temp. Control actual value °Cr
Cooling Temp. Cooling water temperature actual value °Cr
Actual PVW 13 Rotor displacement in the actual value % r
magnetic bearing plane VW13
Actual PVW 24 Rotor displacement in the actual value % r
magnetic bearing plane VW24
Actual PZ 12 Rotor displacement in the actual value % r
magnetic bearing axis Z12
Power Power consumption of the drive actual value W r
Analog Input 1 Analog Input Channel 1 (0...100.0%) actual value % r
Input range 0...10 V
Displ. 0.00...100.0%
Analog Input 2 Analog Input Channel 2 (0...100.0%) actual value % r
Input range 0...10 V
Displ. 0.00...100.0%
* 400 Hz = 24,000 min
-1
for MAG W 830 C
600 Hz = 36,000 min
-1
for MAG W 830, W 1300 (C) & (W) 1500 CT
490 Hz = 29,400 min
-1
for MAG W 2200 (C)
480 Hz = 28,800 min
-1
for MAG W 2800 & 3200 (C/CT)
r (read) = value can only be read
r /w (read/write) = value can be read and written
on (online) = value can be changed always
off (offline) = value can only be changed when the pump stands still
Ready
0.0 A 0 Hz
Enter
Enter
Motor Temp.
50 °C
Enter
Converter Temp.
40 °C
Enter
Bearing Temp.
50 °C
Enter
Actual PVW 13
10%
Enter
Actual PVW 24
10%
Enter
Actual PZ 12
5%
Enter
Power
400 W
Enter
Not Active
Actual Value TMS
70 °C
Not Active
Cooling Temp.
38 °C
Enter
Analog Input 1
30.0%
Enter
T_A
68 °C
Enter
Enter
Analog Input 2
1.5%
Enter
Freq. Setpoint
600 Hz