GA05141_0702 - 09/2004
84
Plug-in control
3.10.5 Menu Set Pump
Menu item Description Adjustable value / option Ac-
cess
min. max. de- Unit
value value fault
Normal Operation Threshold for normal operation
corresponding to the Standby 99 95 % r/w off
frequency setpoint Setpoint
Max. Accel. Time Monitor time for acceleration
and overload 600 3600 1000 s r/w off
Standby Setpoint Speed for Stand-by mode 200 Normal 250 Hz r/w on
= Min. Operation
Freq. Thresh.
3.10.6 Menu Set TMS
Menu item Description Adjustable value / option Ac-
cess
min. max. de- Unit
value value fault
TMS Setpoint Setpoint of the Temperature- 20 80/90* 70 °C r/w on
Management System
* MAG (W) 1500 CT: 90 °C, MAG W 2800, 3200 CT: 80 °C
3.10.7 Menu Set Purge / Vent
Menu item Description Adjustable value / option Ac-
cess
min. max. de- Unit
value value fault
Purge / Vent The operator can select one purge OFF - r/w on
of the beside standing purge ON -
options for purge and vent Controlling purge / vent
via control plug X14 -
Option P / V Displays the factory installed Installed option - r
option for purge and vent
Set Pump
Enter
Normal Operation
95 %
Enter
Max. Accel. time
1000 s
Enter
Set TMS
Enter
TMS Setpoint
70 °C
Enter
Standby Setpoint
250 Hz
Enter
Via X14
Purge ON
Set Purge/Vent
Enter
Purge/Vent
Purge OFF
Enter
Option P/V
Option 1
Enter