GA05141_0702 - 09/2004
75
Plug-in control
3.8 Operation with plug-in control
Observe the general operation rules given in Section 3.1 to 3.5.
Switching on
Switch on the MAG.DRIVE
digital
. The display reads after approximately
20 to 30 seconds
Ready
0.0 A 0 Hz
If the pump has the optional TMS (including e.g. the red heater band) the
heater will be activated. The setpoint temperature will be reached within 30
- 60 minutes depending on cooling water temperature and flow.
In case of corresponding connection the backing pump will be activated
when switching on the MAG.DRIVE
digital
.
Open the purge gas supply.
Open the cooling water supply.
Press the START key.
The pump runs-up.
Acceleration
15.0 A 250 Hz
is diplayed until the frequency setpoint has been reached. Then
Normal Operation
1.0 A 600 Hz
is displayed.