Lighthouse REMOTE-4PN Series Operating Manual
3-12 248083317-1 Rev 4
Data Cable Install
WARNING: Make
sure the data cable is
NOT attached to the hub
before it is connected to
the REMOTE-4PN.
Power to the REMOTE-
4PN should be OFF when
connecting to the hub.
Best practices dictate that power to equipment on both ends of the data
cable should be OFF before a cable is attached. That is not always
practical, especially if the instrument will be attached to a network with
actively-monitored devices and service cannot be interrupted.
If the REMOTE-4PN is involved in a new installation, power should be
OFF at both ends, cables installed and power applied to the REMOTE-
4PN then hub. If power cannot be removed from the hub end, make
sure the REMOTE-4PN is powered OFF before connecting the data
cable.
The required data cable incorporates an O-ring and provides a positive
seal through its twist-lock design. The AC Power cable incorporates
the same twist-lock action. Figures in this section illustrate how both
of these cables are "mated" to the REMOTE-4PN Series.
Note that Figure 3-17 shows the orientation of the key. Make sure the
cable is positioned as shown before connecting. Do NOT force the
connector or it will be damaged.
Figure 3-17 Cable Attachment Step 2
1. Insert the male cable connector into the sensor receptacle, key in
the down position. Make sure the cable fully seats into the
receptacle.