Lighthouse REMOTE-4PN Series Operating Manual
4-16 248083317-1 Rev 4
10. Type telnet xx.xx.xx.xx 9999 and press Enter. This command
accesses the telnet port of the instrument. The expected error
message from Step 9. is shown in Figure 4-14.
Figure 4-14 Telnet 9999
The user will have 4 seconds to respond in the next step. If the user
responds too slowly, they will have to repeat step 10.
11. Press Enter to start the REMOTE-4PN Ethernet interface setup
program. Figure 4-15 displays some of the default settings - make
no changes to these settings except as instructed.
Figure 4-15 Starting the Set Up Program
12. Type 1 to set the IP address. Type the desired IP address and press
Enter. See Figure 4-16.