EasyManua.ls Logo

Lindab LTEST - 10.5 Laboratory Mode; 10.6 Custom Airtightness Class; 10.7 Differential Pressure Measurement

Lindab LTEST
78 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
14
10.5 Laboratory mode
In laboratory mode, the test is carried out without automatic adjustment of the test pressure and without a time limit.
This mode allows the test to be shortened considerably, and is particularly suitable for taking rough measurements.
• After the self test, this screen appears.
• You can press the and keys to configure the pressure and flow rate
manually.
• You can change the adapter during the test – turn down the fan, change
the adapter and press the ADPT key to change the setting.
• Press Stop to end the test. Continue as described above.
10.6 Custom airtightness class
A custom leakage flow rate U can be selected in the start screen in addition to the standardised airtightness classifi-
cations. This means the tests can be performed in other applications that use different classifications, for example in
power plants.
U only appears when the airtightness classes are selected if the value ≠0 is defined.
The standardised air leakage classes are:
A 27 l/s m²
B 9 l/s m²
C 3 l/s m²
D 1 l/s m²
• Press the or key to move to the letter/digit you want to change.
• You can use the and keys to enter a custom leakage flow rate.
• Save the value you entered by pressing the key.
• Alternatively, press the key to exit.
10.7 Differential pressure
In the idle state after it is switched on, the LT 600 can be used as a differential pressure measuring instrument in order
to monitor a pressure curve over time.
The scale is self-scaling, showing a 120 s block that is continuously updated, overwriting the old test curve.
In this mode, you can use the "-" pressure connection (3) if you want to measure the differential pressure between two
test connections rather than using atmospheric pressure. (E.g. for iris diaphragms, filter pressure drops, etc.)
• Press the PD = 0 key to reset the display.
• Press the Back key to return to the menu.
• Press the Stop key to end the test. You can then press Print to print the
graph on the TD 600.
manual-LTEST-1082_EN-DE-FR-SE.indd 14 2012-03-28 12:32:40

Table of Contents

Related product manuals