lion alcolmeter® 600 User Handbook (Ref: RM46914)(ver: 02)
Page 36
Appendix I: Technical Details
Factor
Target
Comments
Error in reading
(active test with a
mouthpiece)
Less than 0.02mg/l when
measuring less than 0.350mg/l
Error in reading
(active test with a
mouthpiece)
Less than 5% when measuring
between 0.350mg/l and
2.000mg/l
Error in reading
(active test with a
mouthpiece)
Less than 20% when
measuring samples of greater
than 2.000mg/l
Error in reading
(active test with a
cup)
Less than 20% or 0.04mg/l
whichever is greater.
Error in reading
(passive test
without a
mouthpiece)
Less than 20% or 0.04mg/l
whichever is greater.
Drift
The test results shall not drift
by more than 5% over a 6
month period