lion alcolmeter® 600 User Handbook (Ref: RM46914)(ver: 02)
Page 37
Clock
The internal clock shall
maintain the correct time and
date to within 5 minutes over
a 12 month period
Frequency
(response and
recovery)
Less than 10
seconds
Samples with 0.000mg/l
Frequency
(response and
recovery)
Less than 20
seconds
Samples of more than
0.000mg/l and 0.100mg/l
Frequency
(response and
recovery)
Less than 30
seconds
Samples of more than
0.100mg/l and 0.250mg/l
Frequency
(response and
recovery)
Less than 60
seconds
Samples of more than
0.250mg/l and 1.000mg/l
Frequency
(response and
recovery)
Less than 120
seconds
Samples of more than
1.000mg/l
Memory effect
Zero
The instruments shall not be
affected by the concentration