lion alcolmeter® 600 User Handbook (Ref: RM46914)(ver: 02)
Page 38
of ethanol presented in the
previous test
Minimum operating
temperature
-5
0
C
Maximum operating
temperature
45
0
C
Operating humidity
range
20% to 95%
non-
condensing
Minimum storage
temperature
-10
0
C
Maximum storage
temperature
60
0
C
Water and dust
ingress
IP-54
Tested and qualified to IP-54
rating
Vibration and shock
Tested in
accordance
with the
specification.
BS EN 60068-2 Test Ea
BS EN 60068-2 Test Fc
EMC testing
Tested in
accordance
with
EN 60601-1-2:2007