lion alcolmeter® 700 User Handbook (Ref: RM79009)(ver: 2)
Warning messages
Fatal error messages
If a fatal message is shown, the lion alcolmeter® 700 automatically powers
down and cannot be used until the appropriate remedial action has been
taken. The most likely occurring examples of such a message are:
Error code messages
The Error code X messages are explained in the following table. If any of
these occur, please contact service personnel for attention.
lion alcolmeter
©
700: Error Code Messages
Fuel cell amplifier signal is too high
Fuel cell output too low during instrument calibration
Gas flow rate applied during breath flow sensor calibration too low
Charge pump did not generate the required voltage
Calibration or checking attempted below 15°C
Calibration or checking attempted above 30°C
Fuel cell offset voltage has drifted too low
Blockage in sampling system
Used in Factory Mode only
Used in Factory Mode only