6. Troubleshooting
Type of Malfunction Possible Cause Correction
- CureJet
TM
not plugged in.
• Plug CureJet
TM
into LED
Controller.
- Unit has exceeded thermal operating limit
(Red indicator light should be on).
• Allow unit to cool.
No light is irradiated from unit when
Start button is depressed.
- Defective Pushbutton. • Call 800-562-8483.
- CureJet
TM
is not operating properly.
Light is irradiated from unit when
Start button is depressed, but blue
indicator light does not illuminate.
- Defective indicator light.
• Call 800-562-8483.
All system functions appear to be
operating, but the product does
not cure completely, or if a radiometer
is used to monitor the light output, the
power is low.
- Lens of CureJet
TM
has an accumulation of
product or other contaminants.
• Clean lens with soft cloth and
isopropyl alcohol.
- LED Controller not plugged in.
• Plug AC line cord into LED
Controller.
- Fuse is blown.
• Replace fuses with 5x20mm 4
Amp fast blow fuses.
Power Switch does not turn unit on.
- Defective Switch. • Call 800-562-8483.