Operator manual Corival
13
4 Specifications
4.1 Technical Specifications
Principle : eddy current brake small disk.
Mains voltage : 115-230 VAC (adjustable)
Frequency : 50 Hz, 60 Hz.
Patient capacity : 350 lb - 160 kg.
Power consumption : 120 VA.
Workload : 0-999 W HYP, 0-999 W LIN
Device control : manual, external analog, external digital
Accuracy : < 3 W for P <= 100 W, < 3% for 100 W<
P<500W, <5% for P>500 W.
Read-out workload : digital, 0-999 resolution 1 W.
Read-out pedalling speed : analog 0-180 rpm or digital 0-255.
Output workload : 100, 200 or 500 W
1 V (max. 999 W).
Output pedalling speed : 100 min
-1
1 V.
Input voltage for external control : 100,200 or 500 W
1 V (max. 999 W).
Safety class Corival : I type B, IEC 601-1 chapter 18, 19 and 20.
Safety class Corival + BP option : IIa type B, IEC 601-1 chapter 18, 19 and 20.
Leakage current : < 0.4 mA.
Isolation resistance : > 4 M
.
Earthway resistance : < 0.2
.
Zero load : < 8 W at 60 rpm.
Noise level : +/- 60dB at workload 200W/60rpm +/-80cm from
crank.
4.2 Fuses
115 V : 2 x 1,60 A slow.
230 V : 2 x 0,80 A slow.
Met opmaak:
opsommingstekens en
nummering

Table of Contents