A Maintenance
The Lode Ergometer is mechanically and electronically designed for durability and accuracy.
This also accounts for the options and add-ons of the device. The frequency of calibration
depends on the intensity of usage. As a rule Lode Ergometers are calibrated on an annual
basis. Please refer to the Service Manual for specific details on check-up intervals for the
Lode Ergometer and its options and/or add-ons.
The dynamic calibration is performed using calibration equipment specially designed by the
manufacturer.
A.1 Cleaning
The housing of the Corival ergometer and all other external surfaces can be cleaned with a
cloth dampened with non-aggressive liquid.
A.2 Replacing fuses
You will not need a certified technician to replace fuses of the Corival.
Always disconnect the power cord first!
Unplug the mains cable. The fuses are part of the main unit (Fig. 3 Main unit panel
connections, item 8) and located at the bottom of the main unit. You have to lift the Corival
place it in the upright position (90°) to locate the fuses. Unscrew the fuse holders with a
screwdriver, replace the broken one and place them back again. Note: make sure that the fuses
that are in the Corival match the set voltage (115 V: 2 x 1,25 A slow; 230 V :2 x 0,63 A
slow).
Contact your local dealer for a maintenance contract or ask Lode:
LODE BV Tel. +31 50 5712811
Service department Fax. +31 50 5716746
Zernikepark 16 Email: ask@lode.nl
9747 AN Groningen Internet: www.lode.nl
The Netherlands