Operator manual Corival
35
The following errors can be displayed:
SNSD
Sensor Disconnect – An absence of signal, meaning the sensor is gone
OOT
Out Of Track – An absence of consecutive good pulse signals
MIDA
SpO2 could not be calculated, missing data indicator
ARTF
Artifact – A detected pulse beat didn’t match the current pulse interval
MPRF
Marginal Perfusion – Amplitude representation of medium signal quality
(holds for entire duration)
LPRF
Low Perfusion – Amplitude representation of low signal quality holds for
entire duration)
TERMINAL
P= 0W 0 rpm
= 55/min SpO2= SNSD
ecs esc