Channel Calibration
MXD70 pH / Redox
Setup and Operating Guide
- 33 -
Channel Calibration
pH Buffer (Offset) + Slope Calibration
Enter the pH Manual Offset and Slope Calibration
Routine
Only available when units set to pH in the channel
setup menu and calibration principle is set to Manual
in this menu.
See page 41 for more details.
/ – Select Option
EXIT
– Return to Select Calibration Channel
– Enter pH Manual Offset Calibration
Offset Value
Displays the electrode Offset currently being used by
the instrument. Only available when units set to pH
in the channel setup menu.
Cannot be edited.
Changed by either using the pH manual offset
calibration, or by the pH auto calibration.
Slope Value
Displays the electrode Slope currently being used by
the instrument. Only available when units set to pH
in the channel setup menu.
Cannot be edited.
Changed by either using the pH manual slope
calibration, or by the pH auto calibration.
pH Buffer (Offset) Calibration
Enter the pH Manual Offset Calibration Routine
Only available when units set to pH in the channel
setup menu and calibration principle is set to Manual
in this menu.
See page 41 for more details.
/ – Select Option
EXIT
– Return to Select Calibration Channel
– Enter pH Manual Offset Calibration