Channel Calibration
- 34 -
MXD70 pH / Redox
Setup and Operating Guide
Channel Calibration
Redox Offset Calibration
The Redox Offset calibration enables the user to
adjust the sensor reading to match a known input.
The current Redox sensor reading can be seen in the
pop-up window and is adjusted by pressing the up
and down arrows. When the reading is correct press
the enter button to store the calibration.
Only available when units set to Redox in the channel
setup menu.
/ – Adjust the Reading Up or Down
EXIT
– Cancel
– Save Calibration
Redox Offset Value
The Redox offset value currently being applied. The
value will change depending on the result of the
Redox offset calibration.
Cannot be edited
Temperature Offset Calibration
The temperature offset calibration enables the user
to adjust the temperature reading to match a known
input. Only available when the channel’s
temperature input is not set to disabled.
The current temperature reading can be seen in the
pop-up window and is adjusted by pressing the up
and down arrows. When the reading is correct press
the enter button to store the calibration.
/ – Adjust the Reading Up or Down
EXIT
– Cancel
– Save Calibration
Temperature Offset Value
The temperature offset value currently being used.
The value will change depending on the result of the
temperature offset calibration.
Cannot be edited