Setpoints
- 18 -
MXD70 Setpoints, Current Outputs, and
Digital Inputs Configuration Guide
Setpoints
Time Remaining
Shows the time remaining till the next clean
operation.
Note – Cannot be edited.
Clean Mode
Associated sensor input channel state when cleaning.
It is recommended that off-line is selected. This will
automatically take the associated sensor input
channel offline, de-energise associated setpoints and
hold associated current outputs, during a clean
operation. This will prevent any undesired control
actions resulting from spraying cleaning solution
onto the sensor.
/ – Select Option
EXIT
– Cancel
– Save Selection
Clean Recovery
If cleaning “Off-line” then the user can introduce an
additional post cleaning delay before coming back
“On-line”, this provides the sensor a period to
stabilise.
/ – Increase / Decrease Digit
– Select Next Digit
EXIT
– Cancel
– Save Value