EasyManuals Logo

LUDLUM 2929 User Manual

LUDLUM 2929
52 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #9 background imageLoading...
Page #9 background image
Model 2929 Scaler Technical Manual Section 2
Ludlum Measurements, Inc. Page 2-1 January 2016
Specifications
0
Power: 95 to 250 Vac wall transformer with 24 Vdc output, 50-60 Hz
single phase (typically less than 100 mA)
Input Senstivity: Beta-gamma lower threshold is adjustable from 4 mV,
beta-gamma upper threshold is adjustable from 25 mV to 100 mV, and
alpha threshold is adjustable from 175 mV.
High Voltage (HV): fully adjustable from 200 to 2500 V ±100 V with
the capability of supporting 60-megohm scintillation loads to 1500 V
HV Meter: 6.4 cm (2.5 in.) panel analog meter readout
HV Meter Accuracy: ±10% of reference value
Audio: click-per-event type with volume control for each of the two
channels
Count Time: 0-99 minutes with multiplier of 0.1, 1.0 and 10; external
position for use with an external clock source
Timer Accuracy: The time base for the scaler is crystal-controlled and
has a timer accuracy of ±0.2% of the thumbswitch setting.
Scaler Readout: six-digit, light-emitting diode readout for each of the
two channels
Scaler Accuracy: ±2% of reference value
Size: 24.4 x 37.1 x 25.1 cm (9.6 x 14.6 x 9.9 in.) (H x W x L)
Weight: 5.5 kg (12.1 lb), including detector
Finish: beige powder coat with anodized front panel and silk-screened
nomenclature
Temperature Range: -20 to 50 °C (-4 to 122°F); may be certified for
operation for -40 to 65 °C (-40 to 150 °F)
Sectio
n
2

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the LUDLUM 2929 and is the answer not in the manual?

LUDLUM 2929 Specifications

General IconGeneral
BrandLUDLUM
Model2929
CategoryMedia Converter
LanguageEnglish