0.1 ℃ -50.1 to -100.0 ℃ ± ( 0.4 % + 1 )℃
-50.0 to 999.9 ℃ ± ( 0.4 % + 0.5 )℃
1 ℃ 1000 to 1300 ℃ ± ( 0.4 % + 1 )℃
0.1 ℉ -58.1 to -148.0 ℉ ± ( 0.4 % + 1.8 )℉
-58.0 to 999.9 ℉ ± ( 0.4 % + 1 )℉
1 ℉ 1000 to 2372 ℉ ± ( 0.4 % + 2 )℉
Type J 0.1 ℃ -50.1 to -100.0 ℃ ± ( 0.4 % + 1 )℃
-50.0 to 999.9 ℃ ± ( 0.4 % + 0.5 )℃
1 ℃ 1000 to 1150 ℃ ± ( 0.4 % + 1 )℃
0.1 ℉ -58.1 to -148.0 ℉ ± ( 0.4 % + 1.8 )℉
** -58.0 to 999.9 ℉ ± ( 0.4 % + 1 )℉
1 ℉ 1000 to 2102 ℉ ± ( 0.4 % + 2 )℉
Type
0.1 ℃ -50.1 to -100.0 ℃ ± ( 0.4 % + 1 )℃
-50.0 to 400.0 ℃ ± ( 0.4 % + 0.5 )℃
0.1 ℉ -58.1 to -148.0 ℉ ± ( 0.4 % + 1.8 )℉
-58.0 to 752.0 ℉ ± ( 0.4 % + 1 )℉
Type E 0.1 ℃ -50.1 to -100.0 ℃ ± ( 0.4 % + 1 )℃
-50.0 to 900.0 ℃ ± ( 0.4 % + 0.5 )℃
0.1 ℉ -58.1 to -148.0 ℉ ± ( 0.4 % + 1.8 )℉
-58.0 to 999.9 ℉ ± ( 0.4 % + 1 )℉
1 ℉ 1000 to 1652 ℉ ± ( 0.4 % + 2 )℉
Type R 1 ℃ 0 to 1700 ℃ ± ( 0.5 % + 3 )℃
1 ℉ 32 to 3092 ℉ ± ( 0.5 % + 5 )℉
Type S 1 ℃ 0 to 1500 ℃ ± ( 0.5 % + 3 )℃
1 ℉ 32 to 2732 ℉ ± ( 0.5 % + 5 )℉
Remark :
a. Accuracy value is specified for the meter only.
b. Accuracy is tested under the meter's environment temperature
within 23 ± 5 .℃
c. Linearity Correction :
Memorize the thermocouple's curve into the intelligent CPU
circuit,
@ Above specification tests under the environment RF Field Strength
less than 3 V/M & frequency less than 30 MHz only.
5