Lutron Electronics YK-2001PHA - 6-2 TDS (PPM) Measurement; 6-3 Conductivity Calibration

Lutron Electronics YK-2001PHA
43 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Change the Temp. Coefficient Factor
The default Temp. compensation factor value of the
measurement solution is to 2.0% per ā„ƒ. If intend to
change it, please refer to chapter 10-7, page 35.
Zero adjustment
If the probe not immerse the measurement solution and
display not show zero value, pressing the " Zero (RANGE) Button "
( 3-5, Fig. 1 ) continuously at least 10 seconds will let
display show zero. The zero function only valid for the
200 uS range and the not zero value is < 2.0 uS.
6-2 TDS ( PPM ) measurement
The measuring procedures are same as above
6-1 Conductivity ( uS, mS ) measurement.
Then use Function key select to TDS Function position.
6-3 Calibration
1) Prepare the standard conductivity solution ( optional )
For example :
2 mS range calibration solution :
1.413 mS Conductivity Standard Solution, CD-14
200 uS range calibration solution :
80 uS Conductivity Standard Solution
20 mS range calibration solution :
12.88 mS Conductivity Standard Solution
or other Conductivity Standard Solution