L98005G3-EU-A 11
User Manual: PCD-51™ 731A
8.3. Storage
1. Keep the console, all components, and garments away from direct
sunlight.
2. Store in a dry, shaded place (see caution below).
3. Do not twist or fold the hose bundle.
4. Store the DC power adapter lightly wrapped and secured.
5. Avoid folding the garments during storage.
Caution:
The console can be transported or stored for short periods of
time within:
Temperature range of -4°F to 158°F (-20°C to 70°C)
Humidity range of 10%rH to 93%rH non-condensing
Atmospheric pressure range of 190hPa to 1060hPa.
Allow the console to reach reasonable room temperature
50°F to 86°F (10°C to 30°C) before operating.
When the system has been stored in extreme temperature
conditions of -20°C (-4°F) or 70°C (158°F) between uses, wait for
two (2) hours before using the system.
8.4. Travel
Make sure to obtain the proper power adapter plug for the location
you visit.
Follow all storage guidelines when packing the console or garment.
8.5. Disposal
If the console, hose bundle and/or garment(s) is to be disposed of,
follow the local country regulations taking environmental factors into