198-003 K Pump IFU Rev D 21
Recommended separation distances between portable and mobile RF
communications equipment and the K Pump.
The K Pump is intended for use in an electromagnetic environment in which radiated RF
disturbances are controlled. The customer r user of the K Pump can help prevent
electromagnetic interference by maintaining a minimum distance between portable and
mobile RF communications equipment (transmitters) and the K Pump as recommended
below, according to the maximum output power of the communications equipment.
Rated maximum
output power of
transmitter
(Watts)
Separation distance according to frequency of transmitter
(Meters)
150 kHz to <80
MHz
80MHz to <800
MHz
800 MHz to 2.5
GHz
𝑑 = 1.17
√
𝑃
𝑑 = 1.17
√
𝑃
𝑑 = 2.33
√
𝑃 800
0,12
0,12
0,12
0,23
0,1
0,37
0,37
0,74
1
0,17
1,17
2,33
10
3,70
3,70
7,37
100
11,7
11,7
23,3
For transmitters rated at a maximum output power not listed above, the
recommended separation distance (d) in meters (m) can be estimated using the
equation applicable to the frequency of the transmitter, where P is the maximum
output power rating of the transmitter in watts (W) according to the transmitter
manufacturer.
Note: These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflections from structures, objects, and people.