198-003 K Pump IFU Rev D 5
Overview & Intended Use
Document Overview
This K Pump Instruction for Use provides information about using the K Pump infiltrator, set-up,
performing cleaning / troubleshooting, and obtaining support.
Product’s Intended Use
The K Pump is designed to infiltrate tumescent fluid into a patient during cosmetic or plastic
surgery procedures.
Electronic Accessories and Equipment
Following is a list of approved electrical accessories for the K Pump:
- Power Supply: Power supply is IEC 60601-1 3
rd
Edition compliant, medical grade; Input
100-240V~1.6a, Max 50-60Hz; Output 24V~2.1a, 50W Max.
- Power Cord: Any six-foot hospital-grade power cord
Optionally if using with M.D. Resource LS system, any 24”
Universal jumper cord
- Foot Pedal: Only use foot pedal supplied by M.D. Resource. To minimize trip hazard,
our recommended maximum length is three meters.
Use of accessories, transducers, and cables other than those sold by Medical Device Resource
may result in system failure, increased EMC emissions, and/or decreased EMC immunity of the
K Pump.
Electromagnetic Interference
This Pump should not be used adjacent to or stacked on other equipment except M.D. Resource
LS2 aspirator pump. If it is used in a stacked configuration, care should be taken to observe
normal operation of system components.
Medical electrical equipment needs special precautions regarding EMC and needs to be installed
and put in service according to instructions and the EMC information included in Appendix C.