198-003 K Pump IFU Rev D 1
K PUMP
Instructions for Use

Questions and Answers

  • D
    derrickbowenAug 4, 2025
    How to troubleshoot a pump that is not operating on M.D. Resource K Pump Medical Equipment?
    • J
      Justin MitchellAug 4, 2025
      If the M.D. Resource Medical Equipment pump isn't operating, first ensure the power switch is turned “ON”. Next, verify the power adapter is properly plugged into both the wall outlet and the unit's power inlet. Finally, make sure to depress the foot pedal to activate the pump.
  • A
    Andrew Gibson Jr.Aug 12, 2025
    What to do if the M.D. Resource K Pump Medical Equipment pump does not start when the foot pedal is depressed?
    • E
      Elizabeth SmithAug 12, 2025
      If the M.D. Resource Medical Equipment pump does not start when the foot pedal is depressed, it could be due to a low fluid flow. Try increasing the speed by turning the control knob clockwise. Also, ensure that the correct tubing is installed in the pump head. Contact M.D. Resource to obtain the proper tubing for the unit.
  • A
    Andrew DuranAug 25, 2025
    Why foot pedals do not activate the pump on M.D. Resource K Pump?
    • D
      David AndersonAug 25, 2025
      If the foot pedals do not activate the M.D. Resource Medical Equipment pump, start by checking the foot pedal cable and ensure it is securely connected to the bulkhead connectors. If the cable is connected, inspect the foot pedal for any damage, such as tears or breaks, and replace the foot pedal if necessary. Also, verify that the power switch is turned “ON”.

Summary

Device Symbols and Markings

General Warning

Symbol indicating a general caution or warning.

Electrical Shock Hazard

Symbol indicating risk of electrical shock.

Overview and Intended Use

Products Intended Use

The specific application of the K Pump in cosmetic surgery.

Safety Information

General Warnings

General warnings and cautions for safe operation.

Safety Guidelines

Essential guidelines for maintaining product quality and safety.

Getting Started

Setup and Operation

Assembly

Steps for assembling the pump, power cord, and footswitch.

Fluid Connection and Priming

Procedure for connecting tubing and priming the pump.

Fluid Delivery

Instructions on adjusting speed and activating fluid flow.

K Pump Features

Connecting Accessories and Devices

Connecting Optional Accessories

Instructions for connecting tubing, IV bag, and cannula.

Activating and Deactivating the K Pump

Activation

Steps to turn on the pump and set it to operate.

Deactivation

Steps to turn off the pump and stop operation.

Cleaning the K Pump

Cleaning the External Case

Instructions for cleaning the exterior of the K Pump.

Support and Warranty Information

Support and Repair

Information on obtaining repair services and RMA procedures.

Product Warranty

Details of the product warranty and limitations.

Troubleshooting

Common Issues and Solutions

Table of common problems and their solutions for the K Pump.

EMC and Interference

Electromagnetic Immunity

Electromagnetic Immunity Guidelines

M.D. Resource K Pump Specifications

General IconGeneral
ManufacturerM.D. Resource
ModelK Pump
Infusion ModesContinuous
Accuracy+/- 5%
Power SourceRechargeable Battery
DisplayLCD
AlarmsOcclusion, Low Battery