Before testing - Preparing the test environment 12
1.2 Preparing the test environment
Make sure that the test environment is as quiet as possible.
Be sure to follow any established infection control procedures for the setting in which you are
working.
Always use clean eartips.
Swab the probe tip with disinfectant between babies.
1.3 Preparing the patient
To prepare the patient for testing, position the patient so that you can easily access the ear to be
tested.
Neonatal screening Neonatal screening is best performed when the baby is sleeping. If required, swaddle newborn
babies before ABR testing for a calming effect.
Grasp the pinna and gently pull back and slightly away from the patient's head.
Look into the ear canal to check for vernix or debris. Clean the ear canal, if required.
Because newborn ear canals are very soft, they are easily pressed out of shape. If this is the case,
wait until the ear canal returns to its original shape.