Otometrics 35
9 Status and error messages
The status and error messages below are listed in alphabetical order.
Error message Cause Solution
Battery charge low. Replace the battery before
continuing with the test.
Cal FAIL Stimulus level calibration
failed.
Ensure stable probe place-
ment; avoid noise.
Change Battery Battery empty < 5.5 Volt. Replace the battery before
continuing with the test.
Clock bat low Clock battery discharged. Contact your service depart-
ment (battery lifetime > 2
years).
Codec failure
Note · Following the message
"Codec failure", the device is
switched off automatically
Codec is defect. Contact your service depart-
ment.