LIMU Write: #LIMU n,m,u set upper limit value and switch on function
n = 1...7 (range)
m = -3000...3000 (digits)
u = 0...2 (unit)
Answer: LIMU n,m,u
Read: ?LIMU
Answer: current stting
LIML Write: #LIML n,m,u set lower limit value and switch on
function
n = 1...7 (range)
m = -3000...3000 (digits)
u = 0...2 (unit)
Answer: LIMU n,m,u
Read: ?LIML
Answer: current setting
SETREL Write: #SETREL n,m,u set relative value and switch on function
n = 1...7 (range),
m = 0...3000 (digits)
u = 0...2 (unit)
Answer: current setting (n,m,u)
Read: ?SETREL
Answer: current setting (n,m,u)
REL Write: #REL 1 on
#REL 0 off
Answer: REL 1 or REL 0
Read: ?REL
Answer: REL 1 or REL 0
PEAK Write: #PEAK 1 on
#PEAK 0 off
Answer: PEAK 1 or PEAK 0
Read: ?PEAK
Answer: PEAK 1 or PEAK 0
Remark: the peak value captured in DC mode is shown in the fist display line
and can be read with ?MEAS.