EasyManua.ls Logo

Maico MA 25 - Page 17

Maico MA 25
40 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Operation Manual MA 25/MA 25e
MA 27/MA 27e
8104818 Rev. 5
16
29/06/2021
Connections
Figure 6
Figure 7
Table 4 Explanation of the Connections
CONNECTIONS
1
Socket for left headphone jack (blue)
2
Socket for right headphone jack (red)
3
Socket for patient response switch
4
Socket for external power supply unit UES18LCPU-050200SPA
4.2.3 Only for MA 25/MA 25e: Battery Compartment
For battery driven use of the MA 25/MA 25e 3x AA batteries have to be placed in the
battery compartment on the backside of the device (Figure 8 and Figure 9).
Figure 8
Figure 9
4.2.4 Establishing a PC-Connection (MA 25e/MA 27e Only)
To transfer data to a PC, establishing a PC-connection via USB is required. If the
MA25e/MA 27e is used with office equipment that is not a medical device itself (see
Table 4, PC-Connection 1), make sure to establish the PC-connection in one of the
following ways (see Table 5, PC Connection 2, 3 or 4).
Make sure you use only office equipment with the device that
is a medical device itself or meets the requirements of
IEC 60950. If a non-medical device is used within the patient
environment (1.5 m from patient as defined in IEC 60601) a
voltage transformer must be used (exception: a battery driven
laptop is used).
WARNING
MA 27/MA27e
MA 25/MA25e
1
1
2
2
3
3
4
4

Other manuals for Maico MA 25

Related product manuals