Operation Manual
touchTymp MI 26
and
MI 36
Version
The status indicator (Figure 40) in the middle of each
test screen provides the probe status on the display
screen.
The same information is shown on the probe with the single LED (Table 11).
Table 11 Test Status Indication
Right ear is selected. Probe is out of ear.
Left ear is selected. Probe is out of ear.
Probe is in the ear and is sealing, test is running or
test is done.
Probe is in the ear and blocked or leaking.
1. Reinsert probe for better placement.
2. Check eartip size and condition.
3. Inspect probe tip for any blockage.
Probe is not attached properly. Check probe
connection.
Probe tone is not given. This status is shortly
shown while the frequency is being changed.
5.4.4 Testing – Immittance
5.4.4.1 Performance and Evaluation of Tympanometry Test
Figure 41 shows the Tympanometry test screen.
NOTE: Tympanometry test screen explanation apply to the Tympanometry module and
the Tympanometry and Acoustic Reflex module.