Site requirements
18
Environmental conditions
Ensure that the following environmental conditions are satisfied. The installation site for
the NanoSampler must be:
l
Indoors - the system is not intended for use outdoors.
l
Well ventilated - especially if noxious samples or dispersants are used.
l
Temperature stable - away from areas where the ambient air temperature could
change, such as air-conditioning, fans or windows.
l
Away from strong light sources - for example, windows allowing in bright sun-
light.
l
Away from strong heat sources - for example, radiators.
The bench that supports the NanoSampler must be:
l
Horizontal
l
Vibration free
l
Capable of supporting the weight of the system.
Electrical considerations
l
Avoid passing electrical cables through areas where liquids can be spilled.
Warning! Do not position the system so that power sockets are obstructed.
They may need to be disconnected during an emergency.
System components weight
Component Weight
NanoSampler Approx. 19kg
Zetasizer Nano (ZSP) 19kg
Computer and Printer See manufacturer’s documentation