Site requirements
19
Operating environment
Operating environment Indoor use only
Operating altitude up to 2000m
Operating temperature 10-40°C
Operating humidity 20-80% RH
Transport and storage requirements
Store the unit in a dry place where no mold will form.
Transport and storage temperature -25-60°C
Transport and storage humidity max. 85% RH
Transport and storage sound pressure level LAeq <70dB
System requirements
The NanoSampler has the following space, power, airline and computer requirements.
Space required
Provide enough space to allow easy access to all components of the system. Allow
2500mm (width) x 650mm (depth) of bench space. This is to allow space for the instru-
ment, and the computer including monitor. Allow at least 800mm above the bench sur-
face for access to the cell area.
Dimensions of the main components of the system are given below.
Component Width Depth Height
NanoSampler
300mm 510mm 360mm
Computer and printer
See manufacturer’s documentation
Power requirements
Warning! Only use the Power Supply Unit (PSU) supplied with the instrument.
Failure to do so could impair operator safety.