Gas Converter CG
®
CG-2/-2M, CG-2H-W/-2MH-W
Instruction Manual
Version 1.04.00

Table of Contents

Questions and Answers

  • R
    robinmitchellAug 7, 2025
    Why is my M&C CG-2 not heating up?
    • E
      Emily MorganAug 7, 2025
      If the M&C Media Converter isn't heating up and LED K1 is off, the heater might be defective, requiring replacement. Alternatively, the temperature controller could be faulty; in this case, measure the voltage at terminal X4/6 and 7. If the voltage is less than 8 V DC, check the controller according to the operating instructions. If it's more than 8 V DC, replace the solid-state relay. A sensor short circuit could also be the cause, so consider changing the sensor. Finally, the solid-state relay itself might be defective and need replacement.
  • N
    Nicholas SmithAug 22, 2025
    What to do if there is no sample gas flow in M&C CG-2 Media Converter?
    • K
      krodriguezAug 22, 2025
      If there's no sample gas flow in your M&C Media Converter, the gas sample lines might be blocked or leaking. Check the gas sample lines and ensure they are gas-tight.

Summary

Safety Instructions

Warranty

Application of Gas Converters

Operating Principle of Conversion

Catalyst Material Selection

Details the available catalyst materials (Type C and Type SS) and their specific requirements.

Technical Data

Installation Instructions

Supply Connections

Electrical Connections

Starting-Up the Converter

Determining Catalyst Service Life

Maintenance

Replacing the Catalyst Cartridge

Step-by-step instructions for safely replacing the catalyst cartridge without tools.

Using the Temperature Controller

Changing Temperature Regulator Parameters

Details how to unlock and modify parameters like SP, ALSE, Lo-t, and rASL for the temperature regulator.

Troubleshooting Guide

M&C CG-2 Specifications

General IconGeneral
CategoryMedia Converter
ModelCG-2
Operating Temperature0°C to 50°C
Fiber TypeSingle-mode or Multi-mode
Power Supply5V DC
Humidity5% to 90% RH (non-condensing)