18 CG-2 | 1.04.00 www.mc-techgroup.com
11 RECEIVING THE CG…
The CG... converter is a completely pre-installed unit. The standard catalyst cartridge supplied is already
installed.
Remove the CG... converter and any accessories carefully from the transport packaging immediately
after arrival and check the contents of the delivery against the delivery note;
Check the goods for any transport damage and - if necessary - inform your transport insurer imme-
diately of any damage.
12 STORAGE
Cartridges should not be stored for more than 1 year. Storage must be vibration-free and dry.
NOTE!
The converter should be stored in a protected frost-free room!
13 INSTALLATION INSTRUCTIONS
The converter housing of the versions CG-2 and CG-2M is designed as a 19" unit and is also suitable
for wall-mounting or mounting plate by using a special mounting bracket. The versions CG-2H-W and
CG-2MH-W are executed as wall mounting housing.
NOTE!
Because of the system of protection IP 20 (EN60529) the converter has to be
built into a protective cabinet respectively housing. When mounting outdoors,
the converter must be installed in a protective housing, frost-free in winter and
adequately ventilated in summer. Direct sunlight is to be avoided.
In order to avoid the re-oxidation of NO to NO
2
in long sample lines, the con-
verter must be mounted in the immediate vicinity of the gas-analyser if possi-
ble.
The converter should be installed away from sources of heat and in a position
which allows air to circulate freely in order to avoid any heat accumulation.
To avoid a heating-up of other devices in case of a 19”-rack, we recommend a
minimum mounting height of 4U.