Section 9: Trouble Shooting
Section 9:
Trouble Shooting
General Purpose Solution:
Set the transmitting device to 1200 BAUD; 8 data bits; no parity. Make sure the data
stream contains 6 weight characters followed by a carriage return, line feed or end of
text. Set the display to factory default and re-learn the display.
The red LED is on and the display reads “NoData”.
Communication was lost.
Suggestions:
Make sure the indicator is powered on.
Make sure the indicator port is enabled to transmit data continuously.
Make sure the wiring is correct. (The corresponding green LED should blink with every
data transmission).
If data delay between data streams is greater than 2 seconds, turn on option 4.
The unit displays the incorrect digits.
Suggestions:
Try shifting the data to the right or left.
Lower the BAUD rate, default the unit, and re-learn
Rice Lake indicators:
Suggestions:
Set End of Line Delay (EOL Delay) to 250 ms or higher.
Do not set to 0 ms.
Display updates slowly.
Suggestions:
Increase the frequency of data transmission.
Turn on option 4.
Default Remote display and re-learn