12
CONSULT
®
HEMOGLOBIN TESTING SYSTEM USER MANUAL
Memory:
Up to 4,000 results with date and time can be stored in the memory of the analyzer. Results from the
memory can be transferred to a PC using the EKF Link software. When the memory of the Consult
®
Hemoglobin analyzer is full, the oldest results will be overwritten consecutively.
Unit setting:
The unit of measurement can be set to g/L, mmol/L or g/dL.
Values are kept in the memory using the unit set at time of measurement.
Bluetooth function:
The Bluetooth function can be activated or deactivated.
Advanced settings
Hematocrit:
Approximate hematocrit value. If this option is activated, the hematocrit value is calculated and displayed
for hemoglobin values between 12.0 and 18.0 g/dL.
If the hemoglobin is inside the normal range, an estimation of the hematocrit is obtained by multiplying
the measured hemoglobin concentration (expressed in g/dL) by factor 2.943. This calculation should not
be used outside the normal range of hemoglobin in humans, e.g. under 12.0 g/dL (7.44 mmol/L) and
above 18.0 g/dL (11.16 mmol/L). It should not be used in anemic conditions. The hematocrit is displayed
for information only and should not be used for clinical decisions.
6.3 DATA TRANSFER
The Consult
®
analyzer comes with a USB 2.0 bus.
For additional information on data transfer, please contact or visit www.ekflink.com for information on EKF
Link software.
6.4 CLEANING AND DISINFECTION
1. Pull the backside of the cuvette holder slightly towards you and lift up.
2. Using a swab, clean the cuvette holder with cold water or a mild detergent, followed by disinfectant.
Dry thoroughly.