the cuvette. If a cuvette cannot be filled in one continuous process, or if the cuvette contains air
bubbles, discard the cuvette and use a new one, repeating steps 2 and 3.
4. Gently wipe off the excess blood on the outside of the cuvette with a gauze pad. Be sure to gently
wipe both sides. Do not wipe too close to the open end as this can draw blood out of the cuvette.
7.3 CONTROL SAMPLING
1. The Consult
®
Hemoglobin Testing System can be verified by use of Consult
®
Hemoglobin Control
Solution. If stored refrigerated, allow the control solution to reach room temperature first. With gloved
hands, take a Consult
®
Hemoglobin Cuvette out of the foil bag and close the bag. Mix the control solution
by gentle inversion 5 times immediately before sampling. Open the vial and discard the first drop.
2. Dispense a second drop of the control solution on to a hydrophobic surface (e.g. Parafilm). Fill the
cuvette completely by touching the corner of the cuvette to the drop. Do not refill the cuvette. If a cuvette
cannot be filled in one continuous process, or if the cuvette contains air bubbles, discard
the cuvette and use a new one with a new drop of control solution.
3. Gently wipe off the excess control solution on the outside of the cuvette with a gauze pad. Be sure
to gently wipe both sides. Do not wipe too close to the open end as this can draw control solution out of
the cuvette.
4. Wipe any excess material from the vial tip and cap with a clean tissue and recap the vial tightly,
immediately.