How to Test Control Solution
Use ONLY McKesson TRUE METRIX®
Control Solution with the McKesson TRUE
METRIX® PRO Professional Monitoring
Blood Glucose Meter and Test Strips.
1. Check dates on control solution label
and test strip vial label. Do not use
control solution or test strips if
Expiration Dates have passed.
(Control solution - 3 months aer
rst opening or date next to EXP
on label; test strips - 4 months aer
rst opening or date next to EXP
on labels.) Discard expired products
and use new products.
2. Allow control solution, vial of test strips and
meter to adjust to room temperature. Write date
first opened on both control solution bottle label
and test strip vial label.
3. Gently swirl or invert control solution bottle to
mix. DO NOT SHAKE!
4. Remove one test strip from vial. Close strip vial
immediately. Use test strip quickly aer removal
from vial.