26
System Out of Range Warning Messages
WARNING!
Meter reads glucose levels from
20-600 mg/dL.
If blood test result is less than 20 mg/dL,
Lo” appears in Display.
If blood test result is greater than 600 mg/dL,
Hi” appears in Display.
A LWAYS repeat test to confirm Low (“Lo”)
and High (“Hi”) results. If results still display
Lo” or “Hi, call your Doctor or Healthcare
Professional immediately.
Note:Lo” results are included in the Average as 20 mg/dL.
Hi” results are included as 600 mg/dL.
If blood glucose test result is greater than
240 mg/dL, and Ketone Test Alert is turned
on, “Ketone” appears in Display with glucose
result (see Ketone Test Alert).
Caution!
When a Ketone Test Alert sounds, it does not
mean that ketones have been detected in the
patient's blood. e Ketone Test Alert is a
reminder to perform a ketone test as prescribed by
the Doctor or Diabetes Healthcare Professional.
Note: Ketone Test Alert can be turned on or o during
Meter Set Up.
Ketone Test
Alert