    Why is weighing not accurate on my Measuretek PS-102-200?
    • J
      jonesmorganAug 13, 2025
      If your Measuretek Scales are not weighing accurately, it could be due to several reasons. First, check for any objects stuck between the load cell and the scale base and remove them. Second, the load cell might have received a heavy impact, in which case you should perform a Linearity calibration. Next, if the scale is not near Chicago, loosen the shipping protection screws. Finally, ensure the load is placed in the center of the weighing platform.
  • F
    Felicia WadeAug 16, 2025
    How to fix Measuretek PS-102-200 that does not turn on?
    • H
      Henry TaylorAug 17, 2025
      If your Measuretek Scales do not turn on, there could be a few reasons: * The AC adapter might not be securely connected. Try re-plugging it or rotating the plug. * The batteries may be low. Replace them with new ones. * The indicator might be damaged. If so, replace it and perform a calibration.
  • L
    Laura PhillipsAug 20, 2025
    What to do if I cannot zero the display on my Measuretek Scales?
    • N
      Nichole MartinAug 20, 2025
      If the display on your Measuretek Scales cannot be zeroed, it could be due to these reasons: 1. The load on the scale might exceed allowable limits. Remove some of the load. 2. The load on the scale might be unstable. Wait for it to stabilize, then press the ZERO/ON/OFF key.
  • F
    fclarkAug 25, 2025
    What does '0 ???' mean on Measuretek PS-102-200 Scales and how to fix it?
    • G
      George ReedAug 25, 2025
      If your Measuretek Scales display shows '0 ???', it might be due to the following reasons: * The indication is out of the key zero range. Reduce the weight on the platform until the indication is within the key zero range. * The weight reading is below the Power On Zero limit. In this case: 1. Install the platform on the scale. 2. Loosen the shipping protection screws. 3. Check for any objects stuck between the scale base and remove them. 4. Perform zero calibration.
  • A
    Audrey LopezAug 27, 2025
    What does 'CAL.Er' mean on Measuretek PS-102-200 Scales and how to resolve it?
    • A
      angelica01Aug 27, 2025
      If your Measuretek Scales display shows 'CAL.Er', it could be due to these reasons: * The input data or loaded weight is either too small or too big. Input correct data and load the correct weight onto the platform. * The weight signal is unstable or un-linear. In this case, return the scale for repair.
  • I
    Isabella MccoyAug 31, 2025
    What does 'EEP.E1' mean on Measuretek PS-102-200 Scales and how can I fix it?
    • K
      Kirk CastilloAug 31, 2025
      If your Measuretek Scales display shows 'EEP.E1', it could be due to: * CONFIG parameters not correctly set. Reset CONFIG parameters per the Technical Manual. * CAL parameters not correctly set. Perform calibration.
  • A
    Angela RomeroSep 4, 2025
    What does 'Ad ???' mean on Measuretek PS-102-200 and how to troubleshoot it?
    • D
      Donna LewisSep 4, 2025
      If your Measuretek Scales display shows 'Ad ???', it could be due to: * The cable from the platform to the indicator is not correctly connected, disconnected, or short circuiting. Replace it with a new indicator and perform calibration. * The indicator is damaged. Return the scale for repair.
  • M
    Matthew WilliamsSep 8, 2025
    What does 'EEP.E2' mean on Measuretek PS-102-200 and how do I fix it?
    • A
      Anna DuncanSep 8, 2025
      If your Measuretek Scales display shows 'EEP.E2', it means the USER parameters are not correctly set. You should reset the USER parameters per the Technical Manual.

Summary

Introduction and Installation

General and Safety Information

Essential safety precautions and general guidelines for operating the scale.

Specifications

Detailed technical specifications of the scale, including capacity and dimensions.

Connection types

Information on the different types of connections available for the scale.

Installation

Step-by-step instructions for properly installing the weighing scale.

Overview of Controls and Functions

Indicator Display

Explanation of the scale's indicator display and character definitions.

Function Keys

Description of the purpose and function of each key on the scale.

Operations

Normal Weighing Mode

Guide on how to perform normal weighing operations with the scale.

Calibration

View ADC Code or Power Voltage

Configuration Parameters Setup

User Parameters Setup

Symbol Definitions

Trouble shooting

Replacement Parts

One Year Limited Warranty

Measuretek PS-102-200 Specifications

General IconGeneral
BrandMeasuretek
ModelPS-102-200
CategoryScales
LanguageEnglish

Related product manuals