What to do if my Medezine Medical Equipment system does not start?
- LLauren AdamsAug 12, 2025
If your Medezine Medical Equipment system isn't starting, several factors could be at play. First, ensure the mains plug fuse is intact and that the power supply fuses are operational. Check that all electrical connectors are properly connected and that the equipment's rated voltage matches your electrical supply. If the filtration unit was used for liquids, disconnect its connector from the power supply. If the yellow overload indicator is on, wait for it to turn off (approximately 20 seconds) before retrying. Confirm that the saw and filtration units are correctly engaged with the power supply. Lastly, if the blade is distorted, index it to use a new section; if blunt, ensure it's correctly fitted per the manufacturer's instructions. If the issue persists, there may be a defective motor.