EQUIPMENT CLASSIFICATION
Class 1 Equipment
Type B Equipment - Protection against electrical shock
Not AP/APG Equipment
Saw Unit Watertight Equipment IP67
Power Supply Splash Proof Equipment IP65
Filtration Unit Drip Proof Equipment IP44
Mode of Operation:-
Continuous with Intermittent Loading up to 3 mins. Duty Cycle 50%
ASSEMBLY INSTRUCTIONS
DO NOT CONNECT TO ELECTRICAL SUPPLY UNTIL ASSEMBLY IS COMPLETED
1 Place chassis on level surface and ensure wing nut is loosened.
2 Lower base of ltration unit into the clamp ring positioning wing nut central about the
power supply.
3 Tighten wing nut.
4 Fit appropriately fused plug to mains input supply lead. This plug is not supplied by
Medezine Ltd.
5 To t blade see page 8.
When only a Saw Unit and Power Supply have been purchased commence assembly
instructions at (4)
SAFETY AND WARNINGS
1 This equipment must be earthed before use.
2 Trained personnel must only use this product.
3 Please read all the instructions before use.
4 Check all equipment before use.
5 Use the appropriate safety equipment.
6 Do not immerse the Power Supply and Vacuum unit in liquid.
7 Use good quality disinfectants as approved by the Hospital Infection Control.
8 Do not use corrosive cleaning materials or Hypochlorite as these will damage the units.
9 Do not leave the Hand Piece soaking in cleaner for prolonged periods of time as will cause
an ingress of uid.