TROUBLESHOOTING- continued
Possible Cause
Saw unit status
indicator lit but not
running although
ltration unit does.
Check the power supply connection
is properly engaged.
Contact manufacturer.
Defective motor.
SolutionSymptom
Faulty connection.
Excessive noise from
saw unit.
Distorted blade.
Badly tting or loose blade.
Filtration unit runs
but does not extract
eciently.
Vacuum hose is kinked
blocked or disconnected.
Repair or replace hosing. For tting
see page 8.
Filter elements not
assembled correctly.
Ensure cloth lter is seated correctly.
Ensure paper lter and support ange
are tted and seated correctly.
Ensure top cover is seated correctly
and toggle clamps are properly
engaged.
Filtration unit does
not extract.
Filtration unit not running.
Check the power supply connection
is properly engaged and the ltration
unit indicator is lit when
the saw unit is running.
Renew blade.
Check blade is tted in accordance
with the manufacturers instructions.
See page 8.
Contact manufacturer.
Renew paper lter, see page 8.
Renew paper lter and check
cloth lter for contamination.
Renew cloth lter, see page 9.
Renew cloth lter and check micron
lter for choking.
Renew micron lter, see page 9.
Paper lter is full.
Paper lter is split.
Cloth lter is choked.
Cloth lter is split.
Micron lter is chocked.
Bearing service required.