13.TechnicalSpecications
Model: BP168A
Measurement Procedure:
Oscillometric, corresponding to Korotko󰀨 method:
Phase I: systolic, Phase V: diastolic
Display: Digital display
Measuring range:
Pressure: 30 to 280 mmHg (in 1 mmHg increment)
Pulse: 40 to 199 beat / minute
Static accuracy:
Pressure: ± 3 mmHg
Pulse: ± 5 % of reading
Measuring resolution: 1 mmHg
Ination: Automatic ination by internal pump
Memory function: 2 x 120 memories for 2 users (SYS, DIA, Pulse)
Decompression: Constant exhaust valve system
Power source 1 (Battery): 4- size “AA” alkaline Batteries
Power source 2 (Adapter):
Input : 100-240V ~ 50 / 60 Hz 0.2A
Output : 6V – 0.6A
Operation temperature: 5 ~ 40°C / 41 ~ 104°F
Operation humidity: 15 % ~ 80 % RH maximum
Storage temperature: -20 ~ +55°C / -4 ~ +131°F
Storage humidity: 10 % ~ 95 % RH maximum
Dimensions: 120 × 80 × 38 ±1.0 mm
Weight: 510 g ± 5g (including batteries and cu󰀨)
Cu󰀨pressuredisplayrange: 0 ~ 290 mmHg / 0 ~ 38.7 kPa
Electrical shock protection: Internal power unit
Safetyclassications:
Type BF equipment
Mode of operation: Continuous operation
Protection against ingress of
water:
IP22
Accessories: M-L size Cu󰀨, 4 “AA” batteries, instruction manual